An independent frequency reference, for checking calibration and measuring carriers.
| Frequency range | Typically 0.1 Hz to 3 GHz with a prescaler |
|---|---|
| Gate time | 0.1 s to 10 s; longer gates give better resolution |
| Timebase | TCXO, typically 0.1-1 ppm stability |
| Resolution | Up to 9 digits with a long gate |
| Sensitivity | Around 20-50 mV RMS for a clean reading |
| Input impedance | 1 Mohm with a selectable 50 ohm input |
| Reference | Internal TCXO, or an external 10 MHz standard |
A frequency counter measures the repetition rate of an input signal by counting cycles over a known gate time. In a station it is used to verify that a transmitter is actually on the frequency it claims, to check a signal generator, or to measure an unknown oscillator.
The accuracy of the measurement depends entirely on the accuracy of the counter's own timebase. A basic crystal reference is good to a few parts per million; a temperature-compensated oscillator is good to a fraction of a part per million; a GPS-disciplined reference is accurate to the point where the counter is no longer the limiting factor.
A counter connected to a transmitter output must be rated for that power level, or connected through a suitable attenuator. Feeding full transmitter power into a counter input destroys it immediately.